Friction-Aire® Safety Oven

Product details

Blue M Friction Aire® ovens provide a controlled heat source without heating elements, which eliminates atmospheric explosions and ignitions when working with hazardous Class 1, Group D materials. Class 1 atmospheres contain flammable vapors, and Group D materials include many solvents commonly used in finishing applications, such as lacquers and paints.

Typical Performance

(Only data differing from typical models is noted under additional models)

SPECIFICATIONS
Model HS-1002, HS-1202, and HS-3802 (Typical) HS-362 HS-1004, HS-1204, and HS-3804
Temperature range 20°C above ambient to 150°C (302°F) -- +20°C above ambient to +260°C (500°F)
Uniformity +100°C: 0.5°C +150°C: 1.0°C +200°C: 1.0°C
Run up time to +150°C: 120 minutes +150°C: 60 minutes +260°C: 140 minutes
Cool down time to +50°C: 30 minutes depending on ambient temperature +50°C: 30 minutes depending on ambient temperature +50°C: 30 minutes depending on ambient temperature
Control accuracy 0.5°C -- --
Approval UL listed Not UL listed Not UL listed
Drift

+50°C: 1.0°C, +100°C: 0.75°C, +150°C: 0.50°C

-- +200°C: 0.50°C
Maximum Velocity 400 / FPM 500 / FPM 500 / FPM
Minimum Velocity 100 / FPM 200 / FPM 250 / FPM
DIMENSIONAL SPECIFICATIONS
Model Inside Dimensions
inches (cm) W x D x H
Overall Dimensions
inches (cm) W x D x H
HP 1Approx Heating Power (Watts) Cubic Feet Capacity Shipping Weight Lbs. (Kg)
HS-1002 14 x 17 x 12
(35 x 43 x 31)
37 x 33 x 48
(93 x 84 x 120)
1 500 1.3 465
(211)
HS-1202 25 x 20 x 20
(63 x 51 x 51)
42 x 39 x 63
(106 x 98 x 159)
2 1300 5.2 800
(363)
HS-3802 38 x 25 x 25
(95 x 63 x 63)
57 x 44 x 68
(144 x 111 x 172)
3 2000 13.7 1300
(590)

1 Voltage - all models: 208/240/480 V 3-Phase / 60Hz. AC Chambers will operate satisfactorily on 208 V 3-Phase / 60Hz. AC when motor is connected for 240 V 3-Phase /60Hz AC

 

DIMENSIONAL SPECIFICATIONS
Model Inside Dimensions
inches (cm)W x D x H
Overall Dimensions
inches (cm) W x D x H
HP 1Approx Heating Power (Watts) Cubic Feet Capacity Shipping Weight Lbs. (Kg)
HS-362-12 37 x 37 x 48
(93 x 93 x 120)
66 x 58 x 91
(167 x 147 x 231)
10 6500 36 2200
(1000)

1 Voltage -all models: 208/240/480 V 3- Phase/ 60Hz. AC Chambers will operate satisfactorily on 208 V 3-Phase / 60Hz. AC when motor is connected for 240 V3-Phase /60Hz. AC

2 Bench Model

 

DIMENSIONAL SPECIFICATIONS
Model Inside Dimensions
inches (cm) W x D x H
Overall Dimensions
inches (cm) W x D x H
HP 1Approx Heating Power (Watts) Cubic Feet Capacity Shipping Weight Lbs. (Kg)
HS-1004-1 14 x 17 x 12
(35 x 43 x 31)
37 x 33 x 48
(93 x 84 x 120)
1 600 1.3 465
(211)
HS-1204-1 25 x 20 x 20
(63 x 51 x 51)
42 x 39 x 63
(106 x 98 x 159)
3 2000 5.2 800
(363)
HS-3804-1 38 x 25 x 25
(95 x 63 x 63)
57 x 44 x 68
(144 x 111 x 172)
5 3000 13.7 1300
(590)

Voltage -all models: 208/240/480 V 3-Phase/ 60Hz AC Chambers will operate satisfactorily on 208 V 3-Phase / 60Hz AC when motor is connected for 240 V 3-Phase /60Hz AC

 

 

   

Request A Quote

Click here to request a quote for one of Blue M's laboratory or industrial ovens.

Blue M Friction Aire® ovens provide a controlled heat source without heating elements, which eliminates atmospheric explosions and ignitions when working with hazardous Class 1, Group D materials. Class 1 atmospheres contain flammable vapors, and Group D materials include many solvents commonly used in finishing applications, such as lacquers and paints.

Typical Performance

(Only data differing from typical models is noted under additional models)

SPECIFICATIONS
Model HS-1002, HS-1202, and HS-3802 (Typical) HS-362 HS-1004, HS-1204, and HS-3804
Temperature range 20°C above ambient to 150°C (302°F) -- +20°C above ambient to +260°C (500°F)
Uniformity +100°C: 0.5°C +150°C: 1.0°C +200°C: 1.0°C
Run up time to +150°C: 120 minutes +150°C: 60 minutes +260°C: 140 minutes
Cool down time to +50°C: 30 minutes depending on ambient temperature +50°C: 30 minutes depending on ambient temperature +50°C: 30 minutes depending on ambient temperature
Control accuracy 0.5°C -- --
Approval UL listed Not UL listed Not UL listed
Drift

+50°C: 1.0°C, +100°C: 0.75°C, +150°C: 0.50°C

-- +200°C: 0.50°C
Maximum Velocity 400 / FPM 500 / FPM 500 / FPM
Minimum Velocity 100 / FPM 200 / FPM 250 / FPM
DIMENSIONAL SPECIFICATIONS
Model Inside Dimensions
inches (cm) W x D x H
Overall Dimensions
inches (cm) W x D x H
HP 1Approx Heating Power (Watts) Cubic Feet Capacity Shipping Weight Lbs. (Kg)
HS-1002 14 x 17 x 12
(35 x 43 x 31)
37 x 33 x 48
(93 x 84 x 120)
1 500 1.3 465
(211)
HS-1202 25 x 20 x 20
(63 x 51 x 51)
42 x 39 x 63
(106 x 98 x 159)
2 1300 5.2 800
(363)
HS-3802 38 x 25 x 25
(95 x 63 x 63)
57 x 44 x 68
(144 x 111 x 172)
3 2000 13.7 1300
(590)

1 Voltage - all models: 208/240/480 V 3-Phase / 60Hz. AC Chambers will operate satisfactorily on 208 V 3-Phase / 60Hz. AC when motor is connected for 240 V 3-Phase /60Hz AC

 

DIMENSIONAL SPECIFICATIONS
Model Inside Dimensions
inches (cm)W x D x H
Overall Dimensions
inches (cm) W x D x H
HP 1Approx Heating Power (Watts) Cubic Feet Capacity Shipping Weight Lbs. (Kg)
HS-362-12 37 x 37 x 48
(93 x 93 x 120)
66 x 58 x 91
(167 x 147 x 231)
10 6500 36 2200
(1000)

1 Voltage -all models: 208/240/480 V 3- Phase/ 60Hz. AC Chambers will operate satisfactorily on 208 V 3-Phase / 60Hz. AC when motor is connected for 240 V3-Phase /60Hz. AC

2 Bench Model

 

DIMENSIONAL SPECIFICATIONS
Model Inside Dimensions
inches (cm) W x D x H
Overall Dimensions
inches (cm) W x D x H
HP 1Approx Heating Power (Watts) Cubic Feet Capacity Shipping Weight Lbs. (Kg)
HS-1004-1 14 x 17 x 12
(35 x 43 x 31)
37 x 33 x 48
(93 x 84 x 120)
1 600 1.3 465
(211)
HS-1204-1 25 x 20 x 20
(63 x 51 x 51)
42 x 39 x 63
(106 x 98 x 159)
3 2000 5.2 800
(363)
HS-3804-1 38 x 25 x 25
(95 x 63 x 63)
57 x 44 x 68
(144 x 111 x 172)
5 3000 13.7 1300
(590)

Voltage -all models: 208/240/480 V 3-Phase/ 60Hz AC Chambers will operate satisfactorily on 208 V 3-Phase / 60Hz AC when motor is connected for 240 V 3-Phase /60Hz AC

 

 

    Features & Benefits

    Features

    • High-velocity airflow system generates heat
    • Blower creates air turbulence and friction for even heat
    • Temperature-sensitive bellows regulates ambient air
    • By-pass damper
    • By-pass chamber

    Benefits

    • Controlled heat source eliminates atmospheric explosions and ignitions when working with Class 1, Group D materials
    • Ideally suited for testing and processing of hazardous materials, paints, solvents, and lacquers.
    • Maximum temperature uniformity
    • Natural temperature interlock and built-in overtemperature protection provides added safety to users and product
    • Economical design that endures years of use

       

    Request A Quote

    Click here to request a quote for one of Blue M's laboratory or industrial ovens.